EasyManua.ls Logo

MBRAUN LABmaster pro sp - Page 163

MBRAUN LABmaster pro sp
302 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 Glovebox or enclosure operation
Setting glovebox parameters
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
8 - 9
The value of the "Upper working limit" must always be higher than the value of the "Lower
working limit".
The value of the "Upper working limit" must at least be 1mbar lower than the "Upper
limit" value.
The value of the "Lower working limit" must at least be 1mbar higher than the "Lower
limit".
The values for "Upper limit" and "Lower limit" are factory set. They can be changed by
MBRAUN Service.
Refer to the chapter Box pressure control [}2 - 19].
Setting of glovebox atmosphere parameters
Start Screen ⇨ Menu ⇨ Box Parameter ⇨ Parameter Boxatmos.
¡ Input of H
2
O and O
2
alarm concentrations
¡ Display of minimum
and maximum H
2
O
and O
2
values since
the last reset
For information on how to
input parameters refer to
section Parameter screens
[}5 - 9].
To reset min. and max.
values:
► Touch the RESET
button.

Table of Contents