8A O2 and H2O analysers
Moisture analyser
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Page
8A - 9
NOTICE
Moisture on the sensor’s metal parts
The metal parts can rust
Moisten only the sensor head if it lies in a horizontal position.
Keep the acid away from all metal parts and the screw joint.
The metal parts can rust.
Moisten only the sensor head (a). The sensor head
must be in horizontal position for moistening.
Keep the acid away from all metal parts and the
screw joint (b).
Apply a thin line of
phosphoric acid (10)
to the sensor (11)
Use a clean, lint-free
cloth to spread the
acid in a thin even
layer around the
sensor (11)
Screw on the protective cap
Screw the protective cap
onto the sensor (12).