2 Inert gas technology
Functional description
Page
2 - 24
MB-20-G/MB-200G-W_MB-LABmaster_MB-200 MOD_TP700
Operating Manual - V5.0 - 04/2018 - STD
Unloading cycle
Glovebox/enclosure Antechamber
1 Antechamber filled with glove-
box/enclosure atmosphere
(equal gas and pressure)
2 Open inner door.
Transfer material into ante-
chamber.
3 Close inner door. 4 Open outer door.
ð Antechamber is filled with
ambient air.
Transfer material to out-
side.
In case of hazardous material or a hazardous atmosphere inside the glovebox, special
safety measures (including e.g. an antechamber cycle) are mandatory. They may not be
part of a standard system.
Antechamber after system downtime
After a lengthy downtime of the system, permeation and diffusion has filled the antecham-
ber with ambient air.
Make a full antechamber cycle (3x evacuate plus refill) prior to start operating the
antechamber.