EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 1013

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SYSTem Subsystem R&S FSL
1300.2519.12 6.254 E-11
S
YSTem:COMMunicate:GPIB[:SELF]:ADDRess
This command changes the GPIB address of the unit.
Parameter
0 to 30
Example
SYST:COMM:GPIB:ADDR 18
Characteristics
*RST value: - (no influence on this parameter, factory default 20)
SCPI: conform
Mode
all
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator
This command changes the GPIB receive terminator.
According to the standard the terminator in ASCII is <LF> and/or <EOI>. For binary data
transfers (e.g. trace data) from the control computer to the instrument, the binary code (0AH)
used for <LF> might be included in the binary data block, and therefore should not be
interpreted as a terminator in this particular case. This can be avoided by changing the receive
terminator to EOI.
Output of binary data from the instrument to the control computer does not require such a
terminator change.
Parameter
LFEOI | EOI
Example
SYST:COMM:GPIB:RTER EOI
Characteristics
*RST value: - (no influence on this parameter, factory default LFEOI)
SCPI: device-specific
Mode
all
SYSTem:COMMunicate:PRINter:ENUMerate[:NEXT?]
This command queries the name of the next printer installed under Windows XP. After all
available printer names have been output, an empty string enclosed by quotation marks (") is
output for the next query. Further queries are answered by a query error.
The SYSTem:COMMunicate:PRINter:ENUMerate:FIRSt? command should be sent
previously to return to the beginning of the printer list and query the name of the first printer.
Example
SYST:COMM:PRIN:ENUM?