TRACe Subsystem R&S FSL
1300.2519.12 6.264 E-11
R
eturn values for Spectrum Emission Mask/Spurious Emissions measurement results
Returns the list evaluation results.
<no>, <start>, <stop>, <rbw>, <freq>, <power abs>, <power rel>, <delta>, <limit check>,
<unused1>, <unused2>
Value Description
<no> range number
<start> start frequency
<stop> stop frequency
<rbw> resolution bandwidth of range
<freq> frequency of peak
<power abs> absolut power in dBm of peak
<power rel> relative power in dBc (related to the channel power) of peak
<delta> distance to the limit line in dB (positive indicates value above the limit, fail)
<limit check> limit fail (pass = 0, fail =1)
<unused1> reserved (0.0)
<unused2> reserved (0.0)
Example
TRAC TRACE1
TRAC? TRACE1
Characteristics
*RST value: –
SCPI: conform
Mode
all
TRACe:COPY
This command copies data from one trace to another.
This command is an event and therefore has no query and no *RST value.
Parameter
TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6,TRACE1 | TRACE2 | TRACE3 |
TRACE4 | TRACE5 | TRACE6
The first operand the destination of the data to be copied, the second operand describes the
source.
Example
TRAC:COPY TRACE1,TRACE2
Characteristics
*RST value: –
SCPI: conform