CALCulate Subsystem (Analog Demodulation, K7) R&S FSL
1300.2519.12 6.290 E-11
CALCulate<1|2>:MARKer:FUNCtion:ADEMod:AM[:RESult<1...6>?]
This command queries the results of the AM modulation measurement. The numeric suffix
(:RESult<1...6>) indicates whether trace 1, 2, 3, 4, 5 or 6 is selected. The numeric suffixes
<1|2> are irrelevant.
Parameter
PPEak Result of measurement with detector +PK
MPEak Result of measurement with detector –PK
MIDDle
Result of averaging ±PK/2
RMS Result of rms measurement
Example
ADEM ON
Switches on the analog demodulator.
CALC:FEED 'XTIM:AM:REL:TDOM'
Switches on the AM result display.
DISP:TRAC ON
Switches on the trace.
CALC:MARK:FUNC:ADEM:AM? PPE
Queries the peak value.
Characteristics
RST value:
SCPI: device–specific
Mode
ADEMOD
CALCulate<1|2>:MARKer:FUNCtion:ADEMod:CARRier[:RESult<1...6>?]
This command queries the carrier power. The numeric suffix (:RESult<1...6>) indicates
whether trace 1, 2, 3, 4, 5 or 6 is selected. The numeric suffixes <1|2> are irrelevant.
With RF Power result display, the carrier power is determined from trace 1 to 6 indicated in the
numeric suffix. With all other result displays, the carrier power is determined from the current
trace data (CLR/WRITE trace).
This command is only a query and therefore has no *RST value.
Example
ADEM ON
Switches on analog demodulator
CALC:FEED 'XTIM:RFP'
Switches on RF power result display
CALC:MARK:FUNC:ADEM:CARR?
Queries the carrier power