SENSe Subsystem (Analog Demodulation, K7) R&S FSL
1300.2519.12 6.306 E-11
ADEM:AM:REL OFF,OFF,OFF
Switches AM demodulation off.
Characteristics
*RST value: OFF,OFF,OFF
SCPI: device–specific
Mode
ADEMOD
[SENSe:]ADEMod:AM:RELative[:TDOMain]:RESult?
This command reads the result data obtained by AM demodulation for the specified result type.
The data format of the output data block is defined by the FORMat command.
Parameter
<result type>: WRITe, AVERage, MAXHold, MINHold; for details see "Trace mode result types"
on page 6.300.
Note: The result type indicated must be one of those configured by
[SENSe:]ADEMod:AM:RELative[:TDOMain][:TYPE]. Otherwise a query error will
be generated.
Return values
ASCII Format (FORMat ASCII) or Binary Format (FORMat REAL,32); for details see "Formats
for returned values: ASCII format and binary format" on page 6.300. The ouput units are
described in "CALCulate<1|2>:MARKer<1...4>:PEXCursion" on page 6.63.
Example
ADEM:SET 8MHz,32000,EXT,POS,–500,30
Sets up demodulator parameters
ADEM:FM AVER,MAXH,MINH
Selects FM results to be measured
ADEM:AM:REL WRIT,OFF,OFF
Selects AM results to be measured
ADEM ON
Switches on demodulator
INIT;*WAI
Starts measurement and waits for sync
FORM ASC
Selects output format
ADEM:FM:RES? AVER
Reads FM average results
ADEM:FM:RES? MAXH
Reads FM max hold results
ADEM:FM:RES? MINH
Reads FM min hold results
ADEM:AM:REL:RES? WRIT
Reads current AM result data
Characteristics
*RST value: –
SCPI: device–specific
Mode
ADEMOD