SENSe Subsystem (Analog Demodulation, K7) R&S FSL
1300.2519.12 6.310 E-11
ADEM:FM:RES? MAXH
Reads FM max hold results
ADEM:FM:RES? MINH
Reads FM min hold results
ADEM:AM:RES? WRIT
Reads current AM results
Characteristics
*RST value: –
SCPI: device–specific
Mode
ADEMOD
[SENSe:]ADEMod:FM:AFSPectrum[:TYPE]
This command selects the AF spectrum result types of the FM demodulated signal to be
measured simultaneously.
Parameter
<result type 1|2|3|4|5|6>: WRITe, AVERage, MAXHold, MINHold, VIEW, OFF; for details see
"Trace mode result types" on page 6.300.
Note: The result type "AF spectrum of the FM demodulated signal" cannot be activated at the
same time as "AF spectrum of AM or PM demodulated signal".
Example
ADEM:FM:AFSP AVER,MAXH,MINH
Determines average, maximum and minimum value simultaneously
ADEM:FM:AFSP WRIT,OFF,OFF
Determines only current measurement results
ADEM:FM:AFSP OFF,OFF,OFF
Switches calculation of AF spectrum off
Characteristics
*RST value: OFF,OFF,OFF
SCPI: device–specific
Mode
A
[SENSe:]ADEMod:FM:AFSPectrum:RESult?
This command reads out the AF spectrum result data of the FM demodulated signal for the
specified result type. The data format of the output data is determined with the FORMat
command.
Parameter
<result type>: WRITe, AVERage, MAXHold, MINHold; for details see "Trace mode result types"
on page 6.300.
Note: The specified result type must be one of those configured with the
[SENSe:]ADEMod:FM:AFSPectrum[:TYPE] command. Otherwise a query error will
be generated.