R&S FSL SENSe Subsystem (Analog Demodulation, K7)
1300.2519.12 6.325 E-11
[SENSe<1|2>:]FILTer:HPASs[:STATE]
This command activates/deactivates the selected high pass filter.
The numeric suffixes <1|2> are irrelevant.
Parameter
ON | OFF
E
xample
FILT:HPAS ON
Activates the selected high pass filter.
Characteristics
RST value: OFF
SCPI: device–specific
Mode
ADEMOD
[SENSe<1|2>:]FILTer:HPASs:FREQuency
This command selects the high pass filter type. For details on filters refer to chapter "Instrument
Functions", section "Analog Demodulation (Option K7)", High Pass AF Filter softkey.
The numeric suffixes <1|2> are irrelevant.
Parameter
50Hz | 300Hz
Example
FILT:HPAS:FREQ 300Hz
Selects the high pass filter for the demodulation bandwidth range from 800 Hz to 16 MHz.
Characteristics
RST value: 300Hz
SCPI: device–specific
Mode
ADEMOD
[SENSe<1|2>:]FILTer:LPASs[:STATE]
This command activates/deactivates the selected low pass filter.
The numeric suffixes <1|2> are irrelevant.
Parameter
ON | OFF
Example
FILT:LPAS ON
Activates the selected low pass filter.
Characteristics
RST value: OFF
SCPI: device–specific
Mode