EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 1131

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SENSe Subsystem (BLUETOOTH, K8) R&S FSL
1300.2519.12 6.372 E-11
Parameter
ON | OFF
Example
ADEM:ZOOM ON
Switches on the zoom function
Characteristics
*RST value: OFF
SCPI: device–specific
Mode
BT
[SENSe:]ADEMod:ZOOM:STARt
The command sets the start time for the display of individual measured values.
The zoom function is enabled or disabled by the [SENSe:]ADEMod:ZOOM[:STATe]
command.
Note: The zoom function is only available for the Output Power, Modulation Characteristics,
Initial Carrier Frequency Tolerance, and Carrier Frequency Drift measurements.
Parameter
0 s to (meas time – 500 / sampling rate)
Example
ADEM:ZOOM ON
Switches on the zoom function
ADEM:ZOOM:STAR 500us
Sets the starting point of the display to 500 µs.
Characteristics
*RST value: 0 s
SCPI: device–specific
Mode
BT

Table of Contents

Related product manuals