R&S FSL CALCulate Subsystem (SPECM, K14)
1300.2519.12 6.403 E-11
Mode
SPECM
CALCulate<1|2>:SPECtrogram:COLor
This command defines the frame color setting.
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
COLor RGB colors
RADar black – green – white
GRAYscale black and white
Example
CALC:SPEC ON
Activates the Spectrogram Measurement option.
CALC:SPEC:COL RAD
Sets the black–green–white color scheme for the frame coloring.
Characteristics
RST value: COL
SCPI: device–specific
Mode
SPECM
CALCulate<1|2>:SPECtrogram:CONTinuous
This command determines whether the results of the last measurement are deleted before
starting a new measurement in single sweep mode.
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
ON | OFF
Example
CALC:SPEC ON
Activates the Spectrogram Measurement option.
INIT:CONT OFF
Sets the single sweep mode.
INIT;*WAI
Starts the measurement and waits for the end of the single sweep measurement.
CALC:SPEC:CONT ON
Repeats the single sweep measurement without deleting the spectrogram results of the last
measurement.
Characteristics
RST value: OFF
SCPI: device–specific
Mode
SPECM