CALCulate Subsystem (CATV, K20) R&S FSL
1300.2519.12 6.430 E-11
CALC:ATV:LIM:RES:HUM:HUM:LOW?
Performs a lower limit check for the hum values.
Characteristics
RST value: –
SCPI: device–specific
Mode
CATV
CALCulate<1|2>:ATV:LIMit:RESult:HUM:HUM:UPPer?
This command performs an upper limit check for the Hum measurement.
The numeric suffixes <1|2> are irrelevant.
This command is a query and therefore has no *RST value.
Example
CALC:ATV:LIM:HUM:HUM:UPP 6
Sets the upper limit of the hum values to 6%.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:ATV:LIM:RES:HUM:HUM:UPP?
Performs an upper limit check for the hum values.
Characteristics
RST value: –
SCPI: device–specific
Mode
CATV
CALCulate<1|2>:ATV:LIMit:RESult:VMODulation:MDEPth:LOWer?
This command performs a lower limit check for the modulation depth of the Vision Modulation
measurement.
The numeric suffixes <1|2> are irrelevant.
This command is a query and therefore has no *RST value.
Example
CALC:ATV:LIM:VMOD:MDEP:LOW 0.5
Sets the lower modulation depth limit of the vision modulation to 0.5%.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:ATV:LIM:RES:VMOD:MDEP:LOW?
Performs a lower limit check for the modulation depth of the Vision Modulation measurement.
Characteristics
RST value: –
SCPI: device–specific
Mode
CATV