CALCulate Subsystem (CATV, K20) R&S FSL
1300.2519.12 6.432 E-11
C
ALCulate<1|2>:ATV:LIMit:RESult:VMODulation:RPC:UPPer?
This command performs an upper limit check for the residual picture in the Vision Modulation
measurement.
The numeric suffixes <1|2> are irrelevant.
This command is a query and therefore has no *RST value.
Example
CALC:ATV:LIM:VMOD:RPC:UPP 80
Sets the upper limit for the residual picture to 80%.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:ATV:LIM:RES:VMOD:RPC:UPP?
Performs an upper limit check for the residual picture in the Vision Modulation measurement.
Characteristics
RST value:
SCPI: device–specific
Mode
CATV
CALCulate<1|2>:ATV:LIMit:RESult:VMODulation:VCPower:LOWer?
This command performs a lower limit check for the vision carrier power in the Vision Modulation
measurement.
The numeric suffixes <1|2> are irrelevant.
This command is a query and therefore has no *RST value.
Example
CALC:ATV:LIM:VMOD:VCP:LOW –50
Sets the lower vision carrier power limit to 10 dBm.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:ATV:LIM:RES:VMOD:VCP:LOW?
Performs a lower limit check for the vision carrier power in the Vision Modulation measurement.
Characteristics
RST value:
SCPI: device–specific
Mode
CATV