CALCulate Subsystem (CATV, K20) R&S FSL
1300.2519.12 6.436 E-11
Example
CONF:ATV:MEAS CSO
Configures the analog TV Carrier–to–Second Order Beat Ratio measurement.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:ATV:RES:CSO? RPOW
Reads the result of the reference power parameter.
or
CALC:ATV:RES:CSO? ALL
Reads all measured values.
Characteristics
RST value: –
SCPI: device–specific
Mode
CATV
CALCulate<1|2>:ATV:RESult:CTB?
This command reads the measurement values of the Carrier–to–Composite Triple Beat Ratio
measurements.
The numeric suffixes <1|2> are irrelevant.
This command is a query only and thus has no *RST value.
Parameter
RPOWer reference power
NFCorrection noise reference bandwidth
MFRelative measurement frequency relative
CTB carrier–to–composite triple beat ratio
ALL Results of all measurements above, separated by commas.
If a measurement result is not valid, a comma will be output only.
Example
CONF:ATV:MEAS CTB
Configures the analog TV Carrier–to–Composite Triple Beat Ratio measurement.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:ATV:RES:CTB? MFR
Reads the result of the relative measurement frequency parameter.
or
CALC:ATV:RES:CTB? ALL
Reads all measured values.
Characteristics
RST value: –
SCPI: device–specific
Mode
CATV