R&S FSL CALCulate Subsystem (CATV, K20)
1300.2519.12 6.437 E-11
CALCulate<1|2>:ATV:RESult:HUM?
This command reads the measurement values of the Hum measurement.
The numeric suffixes <1|2> are irrelevant.
This command is a query only and thus has no *RST value.
Example
CONF:ATV:MEAS HUM
Configures the analog TV Hum measurement.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:ATV:RES:HUM?
Reads the measured value.
Characteristics
RST value: –
SCPI: device–specific
Mode
CATV
CALCulate<1|2>:ATV:RESult:VMODulation?
This command reads the measurement values of the Vision Modulation measurements.
The numeric suffixes <1|2> are irrelevant.
This command is a query only and thus has no *RST value.
Parameter
VCPower vision carrier power
RPC residual picture carrier
MDEPth modulation depth
ALL Results of all measurements above, separated by commas.
If a measurement result is not valid, a comma will be output only.
Example
CONF:ATV:MEAS VMOD
Configures the analog TV Vision Modulation measurement.
INIT;*WAI
Starts a new measurement and waits for the sweep end.
CALC:ATV:RES:VMOD? MDEP
Reads the result of the modulation depth parameter.
or
CALC:ATV:RES:VMOD? ALL
Reads all measured values.
Characteristics
RST value: –
SCPI: device–specific
Mode
CATV