DISPlay Subsystem (CATV, K20) R&S FSL
1300.2519.12 6.478 E-11
D
ISPlay[:WINDow<1|2>]:ZOOM:OVERview
This command sets the zoom for the Overview measurement.
T
he numeric suffixes <1|2> are irrelevant.
Parameter
NONE The whole result table is displayed.
<parameter> The result table of the selected parameter is displayed.
with parameter = MERRms | MERPeak | EVMRms | EVMPeak | CFOFfset
| SROFfset
Example
CONF:DTV:MEAS MERR
Configures the R&S FSL for the Modulation Errors measurement.
DISP:ZOOM:OVER MERP
Displays the measurement results of the peak of the modulation error rate.
Characteristics
RST value: NONE
SCPI: device–specific
Mode
CATV
DISPlay[:WINDow<1|2>]:ZOOM:QUADrant
This command sets the zoom for the constellation diagram.
The numeric suffixes <1|2> are irrelevant.
Parameter
NONE The whole constellation diagram is displayed.
1 | 2 | 3 | 4 The selected quadrant is displayed.
Example
CONF:DTV:MEAS CONS
Configures the R&S FSL for the Constellation Diagram measurement.
DISP:ZOOM:QUAD 1
Display the first quadrant of the constallation diagram.
Characteristics
RST value: NONE
SCPI: device–specific
Mode
CATV