R&S FSL DISPlay Subsystem (Noise Figure, K30)
1300.2519.12 6.537 E-11
D
ISPlay[:WINDow<1>]:TRACe<1|2>:SYMBols
This command activates or deactivates the display of the symbols to mark the measurement
points for the specified trace.
The numeric suffixes <1|2> indicate the measurement results: Trace 1 is always used for noise
results, and trace 2 is always used for gain results.
P
arameter
ON | OFF
Example
DISP:TRAC ON
Switches on the display of trace 1 (noise results).
DISP:TRAC:SYMB ON
Switches on the display of symbols for trace 1 (noise results)
Characteristics
*RST value: OFF for both traces
SCPI: device–specific
Mode
NF
DISPlay[:WINDow<1>]:TRACe<1|2>:X[:SCALe]
This command selects the frequency to be displayed on the x–axis if the DUT is not an
amplifier, i.e. in a frequency–converting measurement mode.
The numeric suffixes <1|2> indicate the measurement results: Trace 1 is always used for noise
results, and trace 2 is always used for gain results.
Parameter
IF | RF
Example
CONF:MODE:DUT DOWN
The DUT converts the input frequency to a lower output frequency.
DISP:TRAC:X IF
The IF frequency is displayed on x–axis.
Characteristics
*RST value: RF
SCPI: device–specific
Mode
NF
DISPlay[:WINDow<1>]:TRACe<1|2>:Y[:SCALe]:AUTO
This command activates or deactivates the automatic scaling of the Y–axis for the specified
trace display. Automatic scaling provides best fit of the Y–axis to the measurement results.
The numeric suffixes <1|2> indicate the measurement results: Trace 1 is always used for noise
results, and trace 2 is always used for gain results.
Parameter
ON | OFF