R&S FSL SENSe Subsystem (Noise Figure, K30)
1300.2519.12 6.557 E-11
C
haracteristics
*RST value: 2 MHz
SCPI: device–specific
M
ode
NF
[SENSe:]FREQuency:STOP
This command specifies the stop frequency for a new frequency measurement list. If this value
is altered, a new frequency list is generated.
Parameter
<numeric value>
Example
FREQ:STOP 700MHZ
Sets the stop frequency for a new list to a value of 700 MHz.
Characteristics
*RST value: 560 MHz
SCPI: conform
Mode
NF
[SENSe:]SWEep:COUNt
This command specifies the number of sweeps over which the measurement results are
averaged. The higher the number of sweeps, the more accurate the results. However, a high
number of sweeps requires a longer measurement time.
Parameter
1 to 32767
Example
SWE:COUN 10
Sets the number of sweeps for averaging to 10.
Characteristics
*RST value: 1
SCPI: conform
Mode
NF
[SENSe:]SWEep:TIME
For details refer to "[SENSe<1|2>:]SWEep:TIME" on page 6.234.