EasyManua.ls Logo

Rohde & Schwarz R&S FSL3 - Page 993

Rohde & Schwarz R&S FSL3
1734 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SENSe Subsystem R&S FSL
1300.2519.12 6.234 E-11
[
SENSe<1|2>:]SWEep:TIME
In analyzer mode, this command defines the sweep time. The available time values vary
depending on the span setting.
If [SENSe<1|2>:]SWEep:TIME is directly programmed, automatic coupling to resolution
b
andwidth and video bandwidth is switched off.
The numeric suffixes <1|2> are not relevant.
Parameter
refer to data sheet
Example
SWE:TIME 10s
Characteristics
*RST value: – (AUTO is set to ON)
SCPI: conform
Mode
A, ADEMOD, NF, CATV, CDMA, EVDO, OFDM, OFDMA/WiBro, SPECM, WLAN
[SENSe<1|2>:]SWEep:TIME:AUTO
This command controls the automatic coupling of the sweep time to the frequency span and
bandwidth settings.
If [SENSe<1|2>:]SWEep:TIME is directly programmed, automatic coupling is switched off.
The numeric suffixes <1|2> are not relevant.
Parameter
ON | OFF
Example
SWE:TIME:AUTO ON
Switches on the coupling to frequency span and bandwidths.
Characteristics
*RST value: ON
SCPI: conform
Mode
A, CATV, CDMA, EVDO, SPECM

Table of Contents

Related product manuals