R&S FSL CALCulate Subsystem (EVDO, K84)
1300.2519.12 6.637 E-11
CALCulate<1|2>:LIMit<1...8>:ESPectrum:VALue
.This command selects one of the following predefined limit lines. As a prerequisite, the
predefined limit lines must be activated using the
CALCulate<1|2>:LIMit<1...8>:ESPectrum:MODE command. This commands supported for
compatibility with the old FSL-K72. It is obsolete and only the
CALCulate<1|2>:LIMit<1...8>:ESPectrum:MODE command should be used. The user can
specify the exclusively used power class by using a value larger or equal to PMIN and smaller
than the PMAX of that Power Class.
The numeric suffixes <1|2> and <1...8> are irrelevant for this command.
Parameter
-200dBm to 200dBm
Example
Power Class 1 -200 dBm P < 28 dBm
Power Class 2 28 dBm P < 33 dBm
Power Class 3 33 dBm P < 200 dBm
P stands for the expected power value.
CONF:CDP:MEAS ESP
Selects the spectrum emission mask measurement.
CALC:LIM:ESP:MODE MAN
Activates the selected predefined limit lines.
CALC:LIM:ESP:VAL 33
Selects the predefined limit line 33 dBm P < 200 dBm.(PowerClass3)
Characteristics
RST value: 0
SCPI: device–specific
Mode
EVDO