R&S FSL CALCulate Subsystem (EVDO, K84)
1300.2519.12 6.639 E-11
C
ALCulate<1|2>:LIMit<1…8>:PVTime:RVALue
This command sets the reference level in dBm for calculating the limit lines Precondition is that
the automatic mode of power calculation is switched off via the commands
CALCulate<1|2>:LIMit<1…8>:PVTime:REFerence ONCE or .
CALCulate<1|2>:LIMit<1…8>:PVTime:REFerence MAN.
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
-200dBm to 200dBm
Example
CALC:LIM:PVT:REF MAN
Manual reference value for limit lines
CALC:LIM:PVTime:RVA -33.5
Set manual reference value to -33.5
Characteristics
RST value: -20dBm
SCPI: device–specific
Mode
EVDO
CALCulate<1|2>:LIMit<1…8>:PVTime:RESTore
This command restores the standard limit lines for the power versus time measurement. All
changes made to the standard limit lines are lost and the state of these limit lines as they left the
factory is restored.
The numeric suffixes <1|2> are irrelevant for this command.
Example
CALC:LIM:PVT:REST
Reset the PVT limit lines to their default setting
Characteristics
RST value: --
SCPI: device–specific
Mode
EVDO