CALCulate:MARKer Subsystem (WiMAX, K92/K93) R&S FSL
1300.2519.12 6.782 E-11
C
ALCulate<1|2>:MARKer<1>:FUNCtion:POWer:RESult:MAXHold?
This command queries the maximum result values of the adjacent channel power
measurement. An ACPR (Adjacent channel power relative) measurement must have previously
been run with more than one sweep to provide the summary data.
For details on the result output refer to
CALCulate<1|2>:MARKer<1>:FUNCtion:POWer:RESult[:CURRent]?
The numeric suffix <1|2> at CALCulate specifies the measurement window.
Example
CALC2:MARK:FUNC:POW:RES:MAXH?
Characteristics
RST value: –
SCPI: device–specific
Mode
OFDM, OFDMA/WiBro
CALCulate<1|2>:MARKer<1>:FUNCtion:ZOOM
This command defines the ratio to be zoomed around the marker 1 in the selected
measurement window. The default value is 1, where the full trace is shown.
This command is only available for the following result displays:
– Constellation vs Carrier
– Constellation vs Symbol
– PVT Full Burst
– PVT Rising / Falling
– Magnitude Capture Buffer
Parameter
1 to 1000000
Example
CALC:MARK:FUNC:ZOOM 2
Zooms 50% in screen A.
CALC:MARK:FUNC:ZOOM 4
Zooms 25% in screen A.
CALC:MARK:FUNC:ZOOM 1
Deactivates zooming in screen A.
Characteristics
RST value: 1
SCPI: device–specific
Mode