SCPI–Specific Error Messages R&S FSL
1300.2519.12 9.4 E-11
Table 9–3 Execution Error – Error on execution of a command; sets bit 4 in the ESR register
Error code Error text in the case of queue poll
Error explanation
–200
Execution error
Error on execution of the command.
–201
Invalid while in local
The command is not executable while the device is in local due to a hard local control.
Example: The device receives a command which would change the rotary knob state, but the device is in
local so the command can not be executed.
–202
Settings lost due to rtl
A setting associated with hard local control was lost when the device changed to LOCS from REMS or to
LWLS from RWLS.
–210
Trigger error
Error on triggering the device.
–211
Trigger ignored
The trigger (GET, *TRG or trigger signal) was ignored because of device timing considerations.
Example: The device was not ready to respond.
–212
Arm ignored
An arming signal was ignored by the device.
–213
Init ignored
Measurement initialization was ignored as another measurement was already in progress.
–214
Trigger deadlock
The trigger source for the initiation of measurement is set to GET and subsequent measurement is
received. The measurement cannot be started until a GET is received, but the GET would cause an
interrupted–error)
–215
Arm deadlock
The trigger source for the initiation of measurement is set to GET and subsequent measurement is
received. The measurement cannot be started until a GET is received, but the GET would cause an
interrupted–error.
–220
Parameter error
The command contains a faulty or invalid parameter.
–221
Settings conflict
There is a conflict between setting of parameter value and instrument state.
–222
Data out of range
The parameter value lies out of the allowed range of the instrument.
–223
Too much data
The command contains too many data.
Example: The instrument does not have sufficient storage space.
–224
Illegal parameter value
The parameter value is invalid.
Example: The text parameter is invalid , TRIGger:SWEep:SOURce TASTe
–230
Data corrupt or stale
The data are incomplete or invalid.
Example: The instrument has aborted a measurement.
–231
Data questionable
The measurement accuracy is suspect.