Setting Traces – TRACE Key R&S FSL
1300.2519.12 4.50 E-11
POWER
Activates linear power averaging.
The power level values are converted into unit Watt prior to averaging. After the averaging, the
data is converted back into its original unit.
Unlike the LIN softkey, the averaging is always done in W.
This softkey is available from firmware version 1.90.
Remote: CALC:MATH:MODE POW