R&S FSL Setting Traces – TRACE Key
1300.2519.12 4.51 E-11
ASCII file export format
The data of the file header consist of three columns, each separated by a semicolon: parameter name;
numeric value; basic unit. The data section starts with the keyword "Trace <n>" (<n> = number of
stored trace), followed by the measured data in one or several columns (depending on measurement)
which are also separated by a semicolon.
File contents: header Description
Type;FSL; Instrument model
Version;5.00; Firmware version
Date;01.Oct 2006; Date of data set storage
Mode;ANALYZER; Instrument mode
Center Freq;55000;Hz Center frequency
Freq Offset;0;Hz Frequency offset
Span;90000;Hz Frequency range (0 Hz in zero span and statistics measurements)
x–Axis;LIN; Scaling of x–axis linear (LIN) or logarithmic (LOG)
Start;10000;Hz
Stop;100000;Hz
Start/stop of the display range.
Unit: Hz for span > 0, s for span = 0, dBm/dB for statistics measurements
Ref Level;–30;dBm Reference level
Level Offset;0;dB Level offset
Ref Position;75;% Position of reference level referred to diagram limits (0% = lower edge)
y–Axis;LOG; Scaling of y–axis linear (LIN) or logarithmic (LOG)
Level Range;100;dB Display range in y direction. Unit: dB with x–axis LOG, % with x–axis LIN
Rf Att;20;dB Input attenuation
RBW;100000;Hz Resolution bandwidth
VBW;30000;Hz Video bandwidth
SWT;0.005;s Sweep time
Trace Mode;AVERAGE; Display mode of trace: CLR/WRITE,AVERAGE,MAXHOLD,MINHOLD
Detector;AUTOPEAK; Detector set: AUTOPEAK,MAXPEAK,MINPEAK,AVERAGE,RMS,SAMPLE,QUASIPEAK
Sweep Count;20; Number of sweeps set
File contents:
data section of the file
Description
Trace 1:;; Selected trace
x–Unit;Hz; Unit of x values: Hz with span > 0; s with span = 0; dBm/dB with statistics measurements
y–Unit;dBm; Unit of y values: dB*/V/A/W depending on the selected unit with y–axis LOG or % with y–
axis LIN
Values; 501; Number of measurement points
10000;–10.3;–15.7
10180;–11.5;–16.9
10360;–12.0;–17.4
...;...;
Measured values: <x value>, <y1>, <y2>; <y2> being available only with detector
AUTOPEAK and containing in this case the smallest of the two measured values for a