R&S FSL Power Measurements – MEAS Key
1300.2519.12 4.95 E-11
The selection of the standard influences the following parameters:
– channel spacing and adjacent–channel spacing
– channel bandwidth, adjacent–channel bandwidth, and type of filtering
– resolution bandwidth
– video bandwidth
– detector
– # of adjacent channels
– trace averaging (switched off)
Remote: CALC:MARK:FUNC:POW:PRES <standard>
CP / ACP Config
Opens a submenu to configure the channel power and adjacent channel power measurement
independently of the predefined standards (for details see also "To set the channel
configuration" on page 4.77 and "Settings of CP / ACP test parameters" on page 4.80).
# of TX Chan (MC–ACP)
Opens an edit dialog box to enter the number of carrier signals to be taken into account in
channel and adjacent–channel power measurements. Values from 1 to 12 are allowed.
Remote: POW:ACH:TXCH:COUN 4
# of Adj Chan
Opens an edit dialog box to enter the number of adjacent channels to be considered in the
adjacent–channel power measurement. Values from 0 to 12 are allowed.
The following measurements are performed depending on the number of the channels:
0 Only the channel powers are measured.
1 The channel powers and the power of the upper and lower adjacent channel are
measured.
2 The channel powers, the power of the upper and lower adjacent channel, and of the
next higher and lower channel (alternate channel 1) are measured.
3 The channel power, the power of the upper and lower adjacent channel, the power of
the next higher and lower channel (alternate channel 1), and of the next but one
higher and lower adjacent channel (alternate channel 2) are measured.
... ...
12 The channel power, the power of the upper and lower adjacent channel, and the
power of the all higher and lower channels (alternate channel 1 to 11) are measured.
Remote: POW:ACH:ACP 1
Channel Settings
Opens a submenu to define the channel settings.