R&S FSL Power Measurements – MEAS Key
1300.2519.12 4.77 E-11
To measure the power in zero span
1. Press the Time Domain Power softkey to activate the power measurement.
The corresponding submenu is displayed.
2. To limit the power evaluation range, switch on the limits (Limits On/Off softkey) and enter the limits
by using the Left Limit and Right Limit softkeys.
3. Select the type of power measurement by using the Mean or RMS softkey. (RMS or mean power),
the settings for max hold and averaging as well as the definition of limits.
4. To calculate and display the peak value, press the Peak softkey.
5. To calculate and display the standard deviation from the mean value, press the Std Dev softkey.
To set the channel configuration
1. Press the CP, ACP, MC–ACP softkey to active channel or adjacent–channel power measurement.
The corresponding submenu is displayed.
2. To use a predefined standard for measurement, press the CP / ACP Standard softkey (for details
on available standards see "Predefined CP / ACP standards" on page 4.79).
3. To configure the parameters independently of the predefined standards, press the CP / ACP
Config softkey (for details see "Settings of CP / ACP test parameters" on page 4.80).
4. To enter the sweep time, press the Sweep Time softkey.
5. To display the whole diagram, press the Full Size Diagram softkey.
6. To adjust the reference level to the measured channel power, press the Adjust Ref Level softkey.
To measure the occupied bandwidth
1. Press the OBW softkey to activate the measurement of the occupied bandwidth (for details see also
"Measurement of occupied bandwidth" on page 4.78).
The corresponding submenu is displayed.
2. Press the % Power Bandwidth softkey to enter the percentage of power.
3. To change the channel bandwidth for the transmission channel, press the Channel Bandwidth
softkey.
4. To optimize the settings for the selected channel configuration, press the Adjust Settings softkey
(for details see also "Settings of CP / ACP test parameters" on page 4.80).
5. To adjust the reference level to the measured total power after the first sweep, press the Adjust
Ref Level softkey.
To measure signal statistics
• To activate and configure the measurement of the amplitude probability distribution (APD), press
the APD softkey (for details refer to chapter "Advanced Measurement Examples", "Amplitude
Distribution Measurement").
The corresponding submenu is displayed.
• To activate and configure the measurement of the complementary cumulative distribution (CCDF),
press the CCDF softkey (for details refer to hapter "Advanced Measurement Examples", "Amplitude
Distribution Measurement").
The corresponding submenu is displayed.