R&S FSL Power Measurements – MEAS Key
1300.2519.12 4.99 E-11
Remote: CALC:LIM:ACP ON
Remote: CALC:LIM:ACP:<adjacent–channel> 0dB,0dB
Remote: CALC:LIM:ACP:<adjacent–channel>:STAT ON
Remote: CALC:LIM:ACP:<adjacent–channel>:ABS –10dBm,–10dBm
Remote: CALC:LIM:ACP:<adjacent–channel>:ABS:STAT ON
For details refer to chapter "Remote Control – Commands", section "CALCulate:LIMit:ACPower
Subsystem".
Power Mode
Opens a submenu to select the power mode.
This softkey is available from firmware version 1.30.
Clear/Write
If this mode is activated, the channel power and the adjacent channel powers are calculated
directly from the current trace (default mode).
This softkey is available from firmware version 1.30.
Remote: CALC:MARK:FUNC:POW:MODE WRIT
Max Hold
If this mode is activated, the power values are calculated from the current trace and compared
with the previous power value using a maximum algorithm. The higher value is retained. If
activated, the enhancement label Pwr Max is displayed.
This softkey is available from firmware version 1.30.
Remote: CALC:MARK:FUNC:POW:MODE MAXH
Select Trace
Opens an edit dialog box to enter the trace number on which the CP/ACP measurement is to be
performed. Only activated traces can be selected (for details on trace modes see "Trace mode
overview" on page 4.40).
Remote: POW:TRAC 2
Adjust Settings
Automatically optimizes all instrument settings for the selected channel configuration (channel
bandwidth, channel spacing) within a specific frequency range (channel bandwidth). The
adjustment is carried out only once. If necessary, the instrument settings can be changed later.
For details on the settings of span, resolution bandwidth, video bandwidth, detector and trace
averaging see "Settings of CP / ACP test parameters" on page 4.80.
Remote: POW:ACH:PRES ACP