Power Measurements – MEAS Key R&S FSL
1300.2519.12 4.98 E-11
A
CP Ref Settings (MC–ACP)
Opens an edit dialog box to select the transmission channel to which the adjacent–channel
relative power values should be referenced.
TX Channel 1 – 12 Selection of one of channels 1 to 12.
Min Power TX Channel The transmission channel with the lowest power is used as a
reference channel.
Max Power TX Channel The transmission channel with the highest power is used as a
reference channel.
Lowest & Highest
Channel
The outer left–hand transmission channel is the reference
channel for the lower adjacent channels, the outer right–hand
transmission channel that for the upper adjacent channels.
Remote: POW:ACH:REF:TXCH:MAN 3
Remote: POW:ACH:REF:TXCH:AUTO MAX
Limit Checking
Opens a submenu to activate and define the limits for the ACP measurement.
This softkey is available from firmware version 1.30.
Limit Chk On/Off
Activates or deactivates the limit check for the ACP measurement.
Remote: CALC:LIM:ACP ON
Remote: CALC:LIM:ACP:ACH:RES?
Remote: CALC:LIM:ACP:ALT:RES?
Edit ACP Limit
Opens the ACP Limits dialog box to define the limits for the ACP measurement.
The following rules apply for the limits:
– A separate limit can be defined for each adjacent channel. The limit applies to both the upper
and the lower adjacent channel.
– A relative and/or absolute limit can be defined. The check of both limit values can be activated
independently.
– The R&S FSL checks adherence to the limits irrespective of whether the limits are absolute or
relative or whether the measurement is carried out with absolute or relative levels. If both limits
are active and if the higher of both limit values is exceeded, the measured value is marked by a
preceding asterisk.