Power Measurements – MEAS Key R&S FSL
1300.2519.12 4.110 E-11
Edit Sweep List/Close Sweep List
Opens/ closes the Sweep List dialog box. Closing the dialog box updates the measurement
results.
For further details refer to "Ranges and range settings" on page 4.81.
This softkey is available from firmware version 1.80.
Insert before Range
Inserts a new range to the left of the currently focussed range. The range numbers of the
currently focused range and all higher ranges are increased accordingly. The maximum number
of ranges is 20.
For further details refer to "Ranges and range settings" on page 4.81.
This softkey is available from firmware version 1.80.
Remote: ESP:RANG3:INS BEF (Spectrum Emission Mask)
Insert after Range
Inserts a new range to the right of the currently focused range. The range numbers of all higher
ranges are increased accordingly. The maximum number of ranges is 20.
For further details refer to "Ranges and range settings" on page 4.81.
This softkey is available from firmware version 1.80.
Remote: ESP:RANG1:INS AFT (Spectrum Emission Mask)
Delete Range
Deletes the currently focused range, if possible. The range numbers are updated accordingly.
For further details refer to "Ranges and range settings" on page 4.81.
This softkey is available from firmware version 1.80.
Remote: ESP:RANG4:DEL (Spectrum Emission Mask)