R&S FSL Power Measurements – MEAS Key
1300.2519.12 4.111 E-11
Edit Reference Range
Opens the Reference Range dialog box to edit the additional settings used for SEM
measurements.
– Peak Power
Measures the highest peak within the reference range.
– Channel Power
Measures the channel power within the reference range (integral bandwidth method).
If the Channel Power reference power type is activated, the dialog box is extended to define
additional settings:
– Tx Bandwidth
Defines the bandwidth used for measuring the channel power:
minimum span
value
span of reference range
– RRC Filter State
Activates or deactivates the use of an RRC filter.
– RRC Filter Settings
Sets the alpha value of the RRC filter. This pane is only available if the RRC filter is activated.
For further details refer to "Ranges and range settings" on page 4.81.
This softkey is available from firmware version 1.80.
Remote: ESP:RTYP PEAK
Remote: ESP:BWID 1MHZ
Remote: ESP:FILT OFF
Remote: ESP:FILT:ALPH 0.5
List Evaluation
Opens a submenu to edit the list evaluation settings.
This softkey is available from firmware version 1.80.