R&S FSL Power Measurements – MEAS Key
1300.2519.12 4.115 E-11
M
eas Start/Stop
Aborts/restarts the current measurement and displays the status:
Start
The measurement is currently running.
Stop
The measurement has been stopped, or, in single sweep mode, the end of the sweep
has been reached.
This softkey is available from firmware version 1.80.
Remote: ABOR
Remote: INIT:ESP (Spectrum Emission Mask)
Remote: INIT:SPUR (Spurious Emissions)
Remote: INIT:CONM (Spurious Emissions)
Spurious Emissions
Opens a submenu to configure the Spurious Emissions measurement.
The Spurious Emissions measurement defines a measurement that monitors unwanted RF
products outside the assigned frequency band generated by an amplifier. The spurious
emissions are usually measured across a wide frequency range. The Spurious Emissions
measurement allows a flexible definition of all parameters.
This softkey is available from firmware version 1.80.
Remote: SWE:MODE LIST
Adjust X–Axis
Adjusts the frequency axis of measurement diagram automatically so that the start frequency
matches the start frequency of the first sweep range, and the stop frequency of the last sweep
range.
This softkey is available from firmware version 1.80.