Power Measurements – MEAS Key R&S FSL
1300.2519.12 4.114 E-11
– Sweep List
Opens the sweep list dialog box. For details see the Sweep List dialog box.
– Add / Remove
Activates or deactivates power classes to be defined. Up to four power classes can be defined.
The number of active power classes affect the availability of the items of the Used Power
Classes dropdown menu.
This softkey is available from firmware version 1.90.
Remote: CALC:LIM:ESP:PCL ON
Remote: CALC:LIM:ESP:PCL MIN <numeric_value>
Remote: CALC:LIM:ESP:PCL:MAX <numeric_value>
Remote: CALC:LIM:ESP:PCL:COUN <numeric_value>
Remote: CALC:LIM:ESP:PCL:LIM ABS
Load Standard
Opens a dialog box to select an XML file which includes the desired standard specification. For
details on the provided XML files refer to "Provided XML files for the Spectrum Emission Mask
measurement" on page 4.82.
This softkey is available from firmware version 1.80.
Remote: ESP:PRES 'WCDMA\3GPP\DL\PowerClass_31_39.xml'
Save As Standard
Opens the Save As Standard dialog box, in which the currently used SEM settings and
parameters can be saved and exported into an *.xml file. Enter the name of the file in the file
name field. For details on the structure and contents of the XML file refer to "Format description
of Spectrum Emission Mask XML files" on page 4.83.
This softkey is available from firmware version 1.90.
Remote: SENS:ESP:STOR "<file name>"
Restore Standard Files
Copies the XML files from the C:\R_S\instr\sem_backup folder to the C:\R_S\instr\sem_std
folder. Files of the same name are overwritten.
This softkey is available from firmware version 1.80.
Remote: ESP:PRES:REST