R&S FSL Cable TV Measurements (Option K20)
1300.2519.12 4.239 E-11
Analog TV Settings
Opens the Analog TV Settings dialog box to modify the modulation standard parameters. The
displayed parameters are derived from the modulation standard referenced in the active channel
table. If no channel table < none > is activated, the default modulation standard is used (for
details see also "Modulation standards" on page 4.210).
For the modulation standard, the changes made via this softkey are only kept as long as a
measurement channel with the same modulation standard is selected. If a measurement
channel with a different modulation standard is selected, or the PRESET key is pressed, the
changes are lost.
If no channel table < none > is activated, the changes of the default modulation standard are
kept as long as the PRESET key is not pressed.
Remote: SET:TV:STAN DK (TV Standard)
Remote: SET:TV:STAN:AUD FM65MONO (Sound System)
Remote: TV:MST:FILT:GDEL FLAT (Group Delay)
Remote: SET:TV:STAN:COL SEC (Color System)
Remote: TRIG:VID:BLIN 16, TRIG:VID:BFI 2, TRIG:VID:BLIN:TYPE CCIR17 (Bar Line)
Remote: TRIG:VID:QLIN 20, TRIG:VID:QFI 2 (Quiet Line)
Remote: DDEM:SBAN NORM (Sideband Position)
Video Scope
Opens a submenu to measure the luminance signal. For this measurement, the TRACE key is
not available.
For further information refer to "Video Scope – analog TV" on page 4.217.
Remote: CONF:ATV:MEAS VSC
Remote: TRAC? TRACE1