Noise Figure Measurements Option (K30) R&S FSL
1300.2519.12 4.260 E-11
Loss Input Constant/Loss Output Constant
Specifies the constant loss value that is used across the entire frequency range. This parameter
is only editable if, in the Selection list, Constant is selected.
Remote: CORR:LOSS:INP:SPOT 10
Remote: CORR:LOSS:OUTP:SPOT 10
Meas Settings
Opens the Measurement Settings dialog box to modify all settings related to the overall
measurement.
Alternatively, the Measurement Settings dialog box is opened as follows:
AMPT key, with focus on the RF Attenuation field
BW key, with focus on the RBW field
Under Calibration, the second stage correction can be activated or deactivated. For details
refer also to "Calibration" on page 4.248.
Under Analyzer Settings, the general settings for the spectrum analyzer concerning the level,
attenuation and bandwidth of the signal to be measured are defined.
The Measurement Settings dialog box contains the following elements:
Calibration 2nd Stage Correction
Analyzer Settings RBW
Sweep Time
Settling Time
Average
RF Attenuation
Automatic Ref Level
Ref Level
Range
Preamplifier