R&S FSL Noise Figure Measurements Option (K30)
1300.2519.12 4.259 E-11
Loss Settings
Opens the LOSS dialog box to take additional losses due to cables or attenuators into account
that are not considered in the calibration.
Under Loss Input Settings, the additional loss between the noise source and the DUT is
defined. Under Loss Output Settings, the additional loss between the DUT and the analyzer is
defined.
Under Loss Input Table or Loss Output Table, the list can contain up to 100 RF/ENR pairs.
The order of the RF values in the list is not important. To modify this table, proceed as described
in "To edit tables" on page 4.248. Loss tables can be saved and recalled at any time via the FILE
key (for details refer to the section "Saving and Recalling Settings Files – FILE Key").
Additionally to the data saved by the basic unit, all loss input & output data entered in the Loss
Settings dialog box are saved.
The Loss Settings dialog box contains the following elements:
Loss Input Settings Selection
Loss Input Constant
Loss Input Table
Loss Output Settings Selection
Loss Output Constant
Loss Output Table
Remote: CORR:LOSS:INP:TABL 1MHz,10,2MHz,12 (Loss Input Table)
Remote: CORR:LOSS:OUTP:TABL 1MHz,10,2MHz,12 (Loss Output Table)
Selection
Defines the used Loss values.
Constant The values specified in the Loss Input Constant/Loss Output Constant fields
are used for all frequencies. The entries of the Loss Input/Output Table are
ignored.
Table The entries of the Loss Input/Output Table provide the basis for the Loss
values. Between these values the R&S FSL uses interpolated values.
Remote: CORR:LOSS:INP:MODE SPOT
Remote: CORR:LOSS:OUTP:MODE SPOT