3GPP Base Station Measurements (Option K72) R&S FSL
1300.2519.12 4.292 E-11
Adjust Ref Lvl
For further details refer to the Adjust Ref Level softkey in the power measurement menu of the
base unit.
ACP
Activates the Adjacent–Channel Power measurement and opens a submenu for configuration.
This measurement is identical to the ACP measurement of the base unit (see CP, ACP, MC–
ACP softkey in the measurement menu of the base unit), but the default settings are in
accordance with the 3GPP specifications:
Parameter Default setting
CP / ACP Standard W–CDMA 3GPP FWD
# of Adj Chan 2
The R&S FSL measures the channel power and the relative power of the adjacent
channels/alternate channels. The results are displayed below the screen.
Remote: CONF:WCDP:MEAS ALCR
Remote: CALC:MARK:FUNC:POWer:RES? ACP
Spectrum Emission Mask
Activates the Spectrum Emission Mask measurement and opens a submenu. This measurement
compares the signal power in defined carrier offset ranges with the maximum values specified
by 3GPP.
The following default settings are used in accordance with the 3GPP specifications:
Parameter Default setting
CP / ACP Standard W–CDMA 3GPP REV
# of Adj Chan 0
Span 25.5 MHz
BW 50 ms