R&S FSL 3GPP Base Station Measurements (Option K72)
1300.2519.12 4.293 E-11
Additional to the graphical result display, the peak list displays the peaks of the frequency
ranges specified by the 3GPP specification. The table contains the following information:
Tx Power power of the transmission channel
Tx Bandwidth bandwidth used to measure the Tx power
Range upper and lower frequency of the evaluated range relative to the center frequency
RBW bandwidth used to measure the evaluated range
Frequency exact frequency where the peak was detected
Level absolute level of the peak
Pwr Rel peak level relative to the Tx power
Delta level relative to the limit.
Positive Values show that the peak exceeds the limit from the specification. The values
Level, Pwr Rel and Delta therefore change their color to red and an asterisk appears beside
the Delta values.
Remote: CONF:WCDP:MEAS ESP
Remote: TRAC? TRACE1
Limit Line Auto
If activated, sets the limit line automatically according to the power determined in the useful
channel. If the measurement is carried out in continuous sweep mode and the channel power
changes from sweep to sweep, this can result in the limit line being continuously redrawn. This
softkey is activated by default.
If deactivated, the limit line settings are defined via the Limit Line Manual softkey.
Remote: CALC:LIM:ESP:MODE AUTO
Limit Line Manual
Opens the Manual Power dialog box to select one of the following predefined limit lines
according to the expected power range: