R&S FSL WiMAX, WiBro Measurements (Options K92/K93)
1300.2519.12 4.445 E-11
Result summary list
If the results are displayed in tabular form (Display Graph/List softkey), the result summary list is
displayed. It shows the overall measurement results and provides limit checking for result values in
accordance with the selected standard. Result values which are within the limit as specified by the
standard are displayed in green. Result values which are outside of the limits specified by the standard
are displayed in red with a '*' to the left. Results which have no limits specified by the standard are
displayed in white. Limit values are displayed in white (not bold) and can be modified, when focused,
via the keypad. Limits are modified for the currently selected modulation scheme. Each modulation
scheme may have its own set of user defined limits. To reset the limit values to the values specified in
the standard, use the lines menu (LINES key).
IEEE 802.16–2004 OFDM
The results displayed in this list are for the entire measurement. If a specific number of bursts have
been requested which requires more than one sweep, the result summary list is updated at the end
of each sweep. The number of bursts measured and the number of bursts requested are displayed
to show the progress through the measurement. The Min / Mean / Max columns show the minimum,
mean or maximum values of the burst results.
Fig. 4-50 Result summary list for IEEE 802.16–2004 OFDM (example)
IEEE 802.16e–2005 OFDMA, WiBro
For these measurement results, the minimum, mean, and maximum is taken over the analyzed
zones of the current capture buffer content.
Two lists are available:
Result Summary of Analyzed Subframes (list 1)
Result Summary of Analyzed Zone / Segment (list 2)