R&S FSL Instrument Setup and Interface Configuration - SETUP Key
1300.2519.12 4.509 E-11
Comb Frequency
Opens a dialog box to set the comb generator frequency for the internal calibration:
C
omb/1
65.8333 MHz / 1
Comb/64
65.8333 MHz / 64
Comb/65
65.8333 MHz / 65
Remote: Sine: DIAG:SERV:INP:PULS OFF
Remote: Comb: DIAG:SERV:INP:PULS ON
Remote: Comb: DIAG:SERV:INP:PRAT COMB64
Reset Password
Deactivates all set passwords.
Remote: SYST:PASS:RES
Selftest
Initiates the self test of the instrument modules to identify a defective module in case of failure .
All modules are checked consecutively and the test result is displayed.
Remote: *TST?
Selftest Results
Opens the Selftest Result dialog box that contains the test results. In case of failure a short
description of the failed test, the defective module, the associated value range and the
corresponding test results are indicated.
Remote: DIAG:SERV:STE:RES?
Password
Opens an edit dialog box to enter the password. This ensures that the service functions are only
used by authorized personnel.
Remote: SYST:PASS "Password"
Service Function
Opens the Service Function dialog box to start special service functions. For further information
refer to the service manual.
Remote: DIAG:SERV:SFUNC <value>,<value>...