Instrument Setup and Interface Configuration - SETUP Key R&S FSL
1300.2519.12 4.508 E-11
System Messages
Opens the System Messages dialog box that displays the generated system messages in the
order of their occurrence. The most recent messages are placed at the top of the list. Messages
that have occurred since the last display of system messages menu are marked with an asterisk
'*'. The following information is available:
No
device-specific error code
Message
brief description of the message
Component
hardware messages: name of the affected module
software messages: name of the affected software
Date/Time
date and time of the occurrence of the message
If the number of error messages exceeds the capacity of the error buffer, Message buffer
overflow is displayed. To delete messages see Clear All Messages softkey.
Remote: SYST:ERR:LIST?
Clear All Messages
Deletes all system messages. The softkey is only available if the System Messages dialog box
is displayed.
Remote: SYST:ERR:CLE:ALL?
Service
Opens a submenu that contains additional functions for maintenance and/or trouble shooting.
NOTICE
Risk of incorrect operation
The service functions are not necessary for normal measurement operation.
However, incorrect use can affect correct operation and / or data integrity of the
R&S FSL. Therefore, many of the functions can only be used after entering a
password. They are described in the instrument service manual.
Input RF/Cal/TG
Selects the input for measurement.
RF Switches the input of the R&S FSL to the input connector (normal position). This is
the default setting.
Cal Switches the RF input of the R&S FSL to the internal calibration source (65.83 MHz)
and activates the data entry of the output level of the calibration source. Possible
values are 0 dB and -30 dB.
TG Switches the RF input of the R&S FSL to the tracking generator output. Only the
models 13, 16 and 28 offer this setting.