R&S FSL CALCulate Subsystem
1300.2519.12 6.19 E-11
C
ALCulate<1|2>:DELTamarker<1...4>:MINimum:LEFT
This command positions the delta marker to the next higher minimum value to the left of the
current value (i.e. descending X values). The corresponding delta marker will be activated first,
if necessary.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
This command is an event and therefore has no *RST value and no query.
Example
CALC:DELT:MIN:LEFT
Sets delta marker 1 to the next higher minimum to the left of the current value.
Characteristics
RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CATV, CDMA, EVDO, SPECM, WCDMA
CALCulate<1|2>:DELTamarker<1...4>:MINimum:NEXT
This command positions the delta marker to the next higher minimum value of the measured
curve. The corresponding delta marker will be activated first, if necessary.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
This command is an event and therefore has no *RST value and no query.
Example
CALC:DELT2:MIN:NEXT
Sets delta marker 2 to the next higher minimum value.
Characteristics
RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CATV, CDMA, EVDO, SPECM, WCDMA
CALCulate<1|2>:DELTamarker<1...4>:MINimum:RIGHt
This command positions the delta marker to the next higher minimum value to the right of the
current value (i.e. ascending X values). The corresponding delta marker will be activated first, if
necessary.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
This command is an event and therefore has no *RST value and no query.
Example
CALC:DELT:MIN:RIGH
Sets delta marker 1 to the next higher minimum value to the right of the current value.