CALCulate Subsystem R&S FSL
1300.2519.12 6.60 E-11
CALCulate<1|2>:MARKer<1...4>:MAXimum:NEXT
This command positions the marker to the next smaller maximum value of the corresponding
trace.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
This command is an event and therefore has no *RST value and no query.
Note: If no next smaller maximum value is found on the trace (level spacing to adjacent
values < peak excursion), an execution error (error code: –200) is produced.
Example
CALC:MARK2:MAX:NEXT
Positions marker 2 to the next lower maximum value.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CATV, CDMA, EVDO, SPECM, WCDMA
CALCulate<1|2>:MARKer<1...4>:MAXimum:RIGHt
This command positions the marker to the next smaller maximum value to the right of the
current value (i.e. in ascending X values) on the corresponding trace.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
This command is an event and therefore has no *RST value and no query.
Note: If no next smaller maximum value is found on the trace (level spacing to adjacent
values < peak excursion), an execution error (error code: –200) is produced.
Example
CALC:MARK2:MAX:RIGH
Positions marker 2 to the next lower maximum value to the right of the current value.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CATV, CDMA, EVDO, SPECM, WCDMA
CALCulate<1|2>:MARKer<1...4>:MINimum[:PEAK]
This command positions the marker to the current minimum value of the corresponding trace.
The corresponding marker is activated first or switched to marker mode, if necessary.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
This command is an event and therefore has no *RST value and no query.
Note: If no minimum value is found on the trace (level spacing to adjacent values < peak