CALCulate Subsystem R&S FSL
1300.2519.12 6.62 E-11
C
haracteristics
*RST value: –
SCPI: device–specific
M
ode
A, ADEMOD, CATV, CDMA, EVDO, SPECM, WCDMA
CALCulate<1|2>:MARKer<1...4>:MINimum:NEXT
This command positions the marker to the next higher minimum value of the corresponding
trace.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
This command is an event and therefore has no *RST value and no query.
Note: If no next higher minimum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: –200) is produced.
Example
CALC:MARK2:MIN
Positions marker 2 to the minimum value of the trace.
CALC:MARK2:MIN:NEXT
Positions marker 2 to the next higher maximum value.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CATV, CDMA, EVDO, SPECM, WCDMA
CALCulate<1|2>:MARKer<1...4>:MINimum:RIGHt
This command positions the marker to the next higher minimum value to the right of the current
value (i.e. in ascending X direction) on the corresponding trace.
For CDMA2000 and 1xEV-DO code domain measurements (option K82 and K84) the numeric
suffix <1|2> selects the measurement screen. In the other modes the numeric suffix <1|2> is
irrelevant.
This command is an event and therefore has no *RST value and no query.
Note: If no next higher minimum value is found on the trace (level spacing to adjacent values
< peak excursion), an execution error (error code: –200) is produced.
Example
CALC:MARK2:MIN
Positions marker 2 to the minimum value of the trace.
CALC:MARK2:MIN:RIGH
Positions marker 2 to the next higher minimum value to the right of the current value.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, ADEMOD, CATV, CDMA, EVDO, SPECM, WCDMA