CALCulate Subsystem R&S FSL
1300.2519.12 6.72 E-11
CALC:MARK:FUNC:Y?
Queries the level of maxima found
CALC:MARK:FUNC:X?
Queries the frequencies (span <> 0) or time (span = 0) of maxima found.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A
CALCulate<1|2>:MARKer<1...4>:FUNCtion:FPEaks:COUNt?
This query reads out the number of maxima found during the search. If no search for maxima
has been performed, 0 is returned.
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.30.
Example
CALC:MARK:FUNC:FPE 3
Searches the 3 highest maxima for trace 1
CALC:MARK:FUNC:FPE:COUN?
Queries the number of maxima found
Characteristics
*RST value: –
SCPI: device–specific
Mode
A
CALCulate<1|2>:MARKer<1...4>:FUNCtion:FPEaks:SORT
This command sets the sort mode for the search for maxima:
X: the maxima are sorted in the list of responses according to increasing X values
Y: the maxima are sorted in the list of responses according to decreasing Y values
The numeric suffixes <1|2> are irrelevant for this command.
This command is available from firmware version 1.30.
Parameter
X | Y
Example
CALC:MARK:FUNC:FPE:SORT Y
Sets the sort mode to decreasing y values
Characteristics
*RST value: –
SCPI: device–specific
Mode
A