CALCulate Subsystem R&S FSL
1300.2519.12 6.94 E-11
C
haracteristics
*RST value: – (= span/10 on activating the function)
SCPI: device–specific
M
ode
A–F
CALCulate<1|2>:MARKer:FUNCtion:STRack:THReshold
This command defines the threshold above which the largest signal is searched for.
The response unit depends on the settings defined with CALCulate<1|2>:UNIT:POWer.
The numeric suffixes <1|2> are not relevant.
Note: The entry of the search bandwidth is only possible if the Signal Track function is
switched on (CALCulate<1|2>:MARKer:FUNCtion:STRack[:STATe]).
Parameter
–330dBm to +30dBm
Example
CALC:MARK:FUNC:STR:THR –50DBM
Sets the threshold for signal tracking to –50 dBm.
Characteristics
*RST value: –120 dBm
SCPI: device–specific
Mode
A–F
CALCulate<1|2>:MARKer:FUNCtion:STRack:TRACe
This command defines the trace on which the largest signal is searched for.
The numeric suffixes <1|2> are not relevant.
Parameter
1 to 6
Example
CALC:MARK:FUNC:STR:TRAC 3
Defines trace 3 as the trace for signal tracking.
Characteristics
*RST value: 1
SCPI: device–specific
Mode
A–F