CALCulate Subsystem R&S FSL
1300.2519.12 6.96 E-11
t
Trace start
Time offset of
first pulse
Measurement
Time
Measurement
Time
Measurement
Time
Period Period
P
Parameter
<time offset of first pulse>, <measurement time>, <period>, < # of pulses to measure>
Example
DISP:WIND:TRAC:Y:RLEV –10dBm
Sets the reference level to 10 dB
INP:ATT 30 dB
Sets the input attenuation to 30 dB
FREQ:CENT 935.2MHz;SPAN 0Hz
Sets the receive frequency to 935.2 MHz and the span to 0 Hz
BAND:RES 1MHz;VID 3MHz
Sets the resolution bandwidth to 1 MHz and the video bandwidth to 3 MHz
DET RMS
Sets the RMS detector
TRIG:SOUR VID;LEV:VID 50 PCT
Selects the trigger source VIDeo and sets the level of the video trigger source to 50 PCT
SWE:TIME 50ms
Sets the sweep time to 50 ms
INIT;*WAI
Starts the measurement with synchronization
CALC:MARK:FUNC:MSUM? 50US,450US,576.9US,8
Queries 8 bursts with an offset of 50 µs, a test time of 450 µs and a period of 576.9 µs
Characteristics
*RST value: –
SCPI: device–specific
Mode
A–T