R&S FSL INITiate Subsystem
1300.2519.12 6.147 E-11
I
NITiate<1|2>:CONMeas
This command continues a stopped measurement at the current position in single sweep mode.
The function is useful especially for trace functions MAXHold, MINHold and AVERage, if the
previous results are not to be cleared with sweep count > 0 or average count > 0 on restarting
the measurement (INIT:IMMediate resets the previous results on restarting the measurement).
The single sweep mode is automatically switched on. Synchronization to the end of the
indicated number of measurements can then be performed with the commands *OPC, *OPC? or
*WAI. In the continuous sweep mode, synchronization to the sweep end is not possible since
the overall measurement "never" ends.
The numeric suffixes <1|2> are irrelevant for this command.
This command is an event and therefore has no *RST value and no query.
Example
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
INIT:CONM;*WAI
Continues the measurement (next 20 sequences) and waits for the end.
Characteristics
*RST value: –
SCPI: device–specific
Mode
A, CATV, CDMA, EVDO
INITiate<1|2>:CONTinuous
This command determines whether the trigger system is continuously initiated (continuous) or
performs single measurements (single).
This setting refers to the sweep sequence (switching between continuous/single sweep).
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
ON | OFF
Example
INIT:CONT OFF
Switches the sequence single sweep.
INIT:CONT ON
Switches the sequence to continuous sweep.
Characteristics
*RST value: ON
SCPI: conform
Mode
all