INITiate Subsystem R&S FSL
1300.2519.12 6.148 E-11
INITiate<1|2>:DISPlay
This command switches the display during a single sweep measurement on (ON) or off (OFF).
The numeric suffixes <1|2> are irrelevant for this command.
Parameter
ON | OFF
E
xample
INIT:CONT OFF
Switches to single sweep mode
INIT:DISP OFF
Sets the display behavior to OFF
INIT;*WAI
Starts the measurement with display switched off.
Characteristics
*RST value: ON
SCPI: device–specific
Mode
A
INITiate<1|2>:ESPectrum
This command starts a Spectrum Emission Mask measurement.
The numeric suffixes <1|2> are not relevant.
This command is an event and therefore has no *RST value and no query.
This command is available from firmware version 1.80.
Example
INIT:ESP
Starts a Spectrum Emission Mask measurement.
Characteristics
RST value: –
SCPI: device–specific
Mode
A