R&S FSL INPut Subsystem
1300.2519.12 6.151 E-11
I
NPut<1|2>:ATTenuation:AUTO
This command automatically couples the input attenuation to the reference level (state ON) or
switches the input attenuation to manual entry (state OFF).
Parameter
ON | OFF
Example
INP:ATT:AUTO ON
Couples the attenuation set on the attenuator to the reference level.
Characteristics
*RST value: ON
SCPI: conform
Mode
all
INPut<1|2>:GAIN:STATe (option RF Preamplifier, B22)
This command switches the preamplifier on or off. The preamplifier only has an effect below
6 GHz.
Parameter
ON | OFF
Example
INP:GAIN:STAT ON
Switches on 20 dB preamplification.
Characteristics
*RST value: OFF
SCPI: conform
Mode
A, NF, CATV, WCDMA, CDMA, EVDO
INPut<1|2>:IMPedance
This command sets the nominal input impedance of the instrument. The set impedance is taken
into account in all level indications of results.
The setting 75 should be selected, if the 50 input impedance is transformed to a higher
impedance using a 75 adapter of the RAZ type (= 25 in series to the input impedance of
the instrument). The correction value in this case is 1.76 dB = 10 log ( 75 / 50).
Parameter
50 | 75
Example
INP:IMP 75